Figure 3.2 – MCS-08 bus system
A basic configuration of type MCS-08 Mult Channel System is shown in Figure 3.2. In this configuration the
gateway module is the master of the internal N-bus and it simultaneously acts as a slave for the external bus.
The gateway connects the external network with the locally installed MCS-08 hardware modules and transmits
commands and responses to and from the external bus. It scans all hardware modules for their status and then
transmits this status information continuously to the external bus controller.
3.2. Specifications
1-channel A/D Converter: Type MCS-08AD
24-bit Delta-Sigma ratiometric with integral analogue and digital filter
0 mV to 18 mV (unipolar) or -18 mV to +18 mV (bipolar), switchable
Up to 800 measurement values per second
Up to 100000 counts (weight value, force, torque) respective
1 million raw counts (unipolar) respective 2 million raw counts (bipolar)
Calibration and Weighing Functions:
Electronic calibration without test weights (eCal) or calibration by test weights
10 step adjustable digital adaptive filter
Tare, zero, auto zero tracking, motion detection, auto-zero at power-up, save
tare at power-off, increased resolution
5 V DC at 58...1200 Ω, max. 100 mA
Up to 6 load cells à 350 Ω or 18 load cells à 1100 Ω in parallel
4- or 6-wire technique, cable length 250 m/mm² for 6-wire connection
By PC software via gateway module, backup data stored on PC
< 4 ms (delay after each read or write command)
10 to 28 VDC, < 200 mA, not galvanically isolated
Environment and Enclosure:
Between -10 °C and +40 °C at maximum 85% RH, non-condensing
Enclosure & protection class
Polyamide, for DIN-rail mounting, protection class IP20
99 x 22.5 x 114.5 mm (L x W x H), weighs approx. 100 g
N-busN-bus
External bus
N-busN-bus
External bus