DAD 141.1 Technical Manual, Rev. 2.2 - February 2014
7.4. Menu 2 – System Span
Remark: Activate a new calibration with 1x Power OFF/ON !
SPAN setup (Menu 2.1 to 2.4) TAC protected – see chapter 4.12
Set SPAN Calibration value (command CG)
Set display value equivalent to calibration weight or to mV/V value
derived from load cell(s) test data
Calibrate system SPAN - gravimetric by weight / load
Display shows the actual input signal in mV/V
Apply test weight equivalent to calibration value (2.1)
Press ENTER button to store new SPAN signal.
Calibrate system SPAN - electronic by mV/V input (command AG)
Use the UP/DOWN & MOVE RIGHT keys to setup the mV/V value at
which the device should read SPAN
Press ENTER button to store new SPAN signal.
Display the input signal in mV/V
This function displays the actual input signal
7.5. Menu 3 – Display
Remark: Activate a new calibration with 1x Power OFF/ON !
Display setup (Menu 3.1 to 3.3) TAC protected – see chapter 4.12
Display limits - Overrange / Underrange (commands CM/CI)
Display overrange limit (maximum value +999999)
Use the UP/DOWN & MOVE RIGHT keys to setup the maximum
display value, above which the display shows over range (all
dashes in the top of the display).
Display underrange limit (minimum value -999999)
Use the UP/DOWN & MOVE RIGHT keys to setup the minimum
display value, above which the display shows under range (all
dashes in the bottom of the display).
Display step size - in digits [d] (command DS)
choose one out of 1, 2, 5, 10, 20, 50, 100, 200, 500
Decimal point position on the display (command DP)
choose one out of 0, 0.0, 0.00, 0.000, 0.0000, 0.00000