10.1.5. SR Reset DAD 141.1 Firmware
Slave (DAD 141.1) responds
This command will respond with ‘OK’ and after maximum 400 ms perform a complete reset of the DAD 141.1.
It has the same functionality as power OFF and ON again.
10.1.6. RS Read Serial Number [ Index 0x2034 ]
Issuing the RS command will return the current serial number in the format S+12345678.
Slave (DAD 141.1) responds
10.2. Calibration Commands – CE, CM, CI, DS, DP, CZ, CG, ZT, FD, ZR, ZI,
AZ, AG, CS, SU, RU
10.2.1. CE Read TAC* Counter / Open Calibration Sequence [ Index 0x2204 ]
With this command you can read the TAC counter (*TAC = Traceable Access Code) or you can open a
calibration sequence.
Slave (DAD 141.1) responds
Request: TAC counter CE17
Calibration sequence active
This command must be issued PRIOR to any attempt to set the calibration parameters AG, AZ, CM, CI, DS,
DP, CZ, CG, ZT, ZR, FD or CS. In legal for trade applications the TAC counter can be used to check if critical
parameters have been changed without re-verification. After each calibration the TAC counter increases by 1.
10.2.2. CM Set Maximum Output Value [ Index 0x220C ]
This command is used to set up the maximum output value. Permitted values are from 1 to 999 999.
Slave (DAD 141.1) responds
Request: TAC counter CE17
Calibration sequence active
This value will determine the point at which the output will change to “ooooooo”, signifying over-range.
Note: The range, in which a scale can be set to zero (SZ) or automatic zero tracking (ZT) is active, is as
standard +/- 2% of CM value. But with the settings of ZT (see 10.2.8) and/or ZR (see 10.2.10) you can change
the behaviour.
Factory default: CM = 010009
10.2.3. CI Set Minimum Output Value [ Index 0x220E ]
This command is used to set up the minimum output value. Permitted values are from – 999 999 to 0.
Slave (DAD 141.1) responds
Request: TAC counter CE17
Calibration sequence active
This value will determine the point at which the output will change to “uuuuuuu”, signifying under-range.
Note: In bipolar applications (e.g. force- or torque measurements) this parameter defines the max. output value
for input signals with negative sign.
Factory default: CI = –010009